MDP600 SC Sealed Medical Grade Power Supplies

The MDP600 SC Series of medical grade AC/DC power supplies come in an environmentally sealed and potted enclosure that relies on convection and/or conduction cooling for thermal management. By eliminating the fan, you eliminate unwanted by-products such as poor ingress protection, increased acoustical noise, compromised room air antiseptics, and reduced power supply life time, all critical to medical environments.

Key features

  • IP66 / 67 / 68 rated sealed enclosure
    • IP68 characterized as 1m depth for 40 days
  • High 94% operating efficiency
    • 50% load and above
  • Low stand-by power consumption<0.3W
  • Wide range input 90 – 264 VAC
  • 600 W free air cooled, 800 W peak (10s)
  • Single output 24 and 48 VDC variants
  • Low earth/touch leakage currents
    • <300uA
  • Input inrush current limiting
  • 2x Mopp grade protection – BF apparatus compatible

Applications

  • MRI systems / full-body tomography
  • Surgery room operating tables
  • Dental surgery chair
  • X-ray medical apparatus
  • Clinical / laboratory table-top / floor-based analyzers
  • Other medical electronics
Model number
Power Out Max (W)
Input Voltage (VAC)
V1 Out Nom (V)
Output Current Max (A)
Cooling
MDP600-US24-SC
Power Out Max (W) 600
Input Voltage (VAC) 90 - 264
V1 Out Nom (V) 24
Output Current Max (A) 25
Cooling Natural Convection
Max # Outputs
1
Efficiency Rating (%)
94
Operating Altitude (m)
4000
Operating temperature range
-30 to +70°C with de-rating
Safety
IEC/EN 60601-1 3rd ed.; 2x MoPP - BF compatible: IEC/EN 60601-1-2 4th ed. EMC immunity
IP Rating
IP66/67/68
MDP600-US24-SC-S
Power Out Max (W) 600
Input Voltage (VAC) 90 - 264
V1 Out Nom (V) 24
Output Current Max (A) 25
Cooling Natural Convection
Max # Outputs
1
Efficiency Rating (%)
94
Operating Altitude (m)
4000
Operating temperature range
-30 to +70°C with de-rating
Safety
IEC/EN 60601-1 3rd ed.; 2x MoPP - BF compatible: IEC/EN 60601-1-2 4th ed. EMC immunity
IP Rating
IP66/67/68
MDP600-US48-SC
Power Out Max (W) 600
Input Voltage (VAC) 90 - 264
V1 Out Nom (V) 48
Output Current Max (A) 12.5
Cooling Natural Convection
Max # Outputs
1
Efficiency Rating (%)
94
Operating Altitude (m)
4000
Operating temperature range
-30 to +70°C with de-rating
Safety
IEC/EN 60601-1 3rd ed.; 2x MoPP - BF compatible: IEC/EN 60601-1-2 4th ed. EMC immunity
IP Rating
IP66/67/68
MDP600-US48-SC-S
Power Out Max (W) 600
Input Voltage (VAC) 90 - 264
V1 Out Nom (V) 48
Output Current Max (A) 12.5
Cooling Natural Convection
Max # Outputs
1
Efficiency Rating (%)
94
Operating Altitude (m)
4000
Operating temperature range
-30 to +70°C with de-rating
Safety
IEC/EN 60601-1 3rd ed.; 2x MoPP - BF compatible: IEC/EN 60601-1-2 4th ed. EMC immunity
IP Rating
IP66/67/68

Thanks to the combination of the potted PCB and finned aluminum case, the series delivers a steady 600 W (800 W peak for 10s) of regulated DC power up to the 55°C ambient temperature, in a compact 4.92 x 9.86 x 2.36″ form factor, facilitating optimal thermal management in space constrained environments with very high reliability.

The input/output connections utilize flying leads fixed to the chassis through watertight glands which give the power supply an IP66/67/68 ingress protection rating (IP68 characterized as 1m depth for 40 days), suitable for harsh environments. The MDP600 SC series offers high 94% typical conversion efficiency, low stand-by power consumption of less than 0.3 W, and 0.9 power factor, ensuring compliance with the latest DOE and ErP directives for energy saving in medical equipment.

The MDP600 SC standard version is available in single 24 and 48 VDC main output voltages. It is also available in the (-S) variant which offers an additional +5V stand-by output and the following control signals: +/- remote sense, remote On/Off (PS_inhibit), Power Good (PS_OK), and I-share (ISHARE1+V_SLOGIC). Additional protection features include high breaking capacity fuses on both AC lines, input under voltage lockout (IUV), output over-current (OC), output short-circuit (SC), output over-voltage (OV), and over-temperature (OT).

The MDP600 SC series is approved to the 3rd edition of the ANSI/AAMI ES60601-1, IEC/EN 60601-1 standards for medical grade power supplies, including 2x MoPP means of patient protection and BF appliances compatibility. The series also complies with the EN 55011, EN 60601-1-2 Class B limits for conducted and radiated emissions, IEC/EN 61000-3 for harmonic and flicker, and IEC/EN 60101-1-2 4th edition for EMC immunity standards.